The ATxIsoBlock is a 2-wire in-head temperature transmitter with galvanic isolation between input and output, microprocessor based designed for the flexibility of accepting mV, Pt100 and a variety of thermocouples as the input sensor. The ATxIsoBlock delivers a scalable linear 4-20mA output current proportional to the sensor temperature. User-friendly configuration software is provided for parameter setup, including sensor type, temperature range, etc. |
|
Sensor Input | : | User defined. The supported sensors are listed in ATxIsoBlock Catalogue, Input Sensors, along with their maximum ranges.
| Thermocouples | : | Types J, K, R, S, T, N, B and E according to IEC 60584. Impedance: 1 MΩ. | Pt100 | : | Excitation: 180 µA. 2 or 3-wire connection ( for 2-wire sensors, tie terminals 2 or 3 together ). α = 0.00385, according to IEC 60751. | Voltage | : | 0 ~ 50 mV DC. Impedance: 1 MΩ. | Total Accuracy | : | Better than 0.25% of the maximum range for Thermocouples and 0.15% for Pt100 and Voltage. | Response Time | : | < 100 ms. | Output | : | 2-wire 4-20 mA, linear with respect to the measured temperature. | Resolution | : | 0.001 mA (14 bits). | Power Supply | : | 12 to 35 Vdc, across the transmitter. | Maximum Load (RL) | : | RL (max.)= (Vcc – 12) / 0.02 Ω. Where: Vcc= Power supply voltage. | Operating Temperature | : | -40 to 85 °C. | Humidity | : | 0 to 90% RH. | Electromagnetic compatibility | : | EN 50081-2, EN 50082-2. | Polarity Inversion | : | Internal Protection. | Isolation | : | 1500 Vac isolation for 1 minute between the sensor and the 4-20mA loop. | Misc. | : | Cold junction compensation for thermocouples. | Housing | : | ABS plastic. | Dimensions | : | 44mm (Diameter) x 25mm (Height). |
|
|
[ATxISOBlock_Universal_Temperature_Transmitter.pdf]
|
[ATxConfig_ABUSTEK_Software.zip]
|
|
|
|